Designed for reliable and eco-conscious laboratory storage, this 30 cu. ft. touchpad laboratory refrigerator offers versatile performance for lab environments. Its temperature control is powered by microprocessor technology, and maintains tight temperature uniformity with an adjustable operating range of 2°C to 10°C. Enjoy a user-friendly touchpad controller and a robust alarm system including audible and visual temperature alarms, door ajar alarms, sensor error alarms, and power failure alarms—all with battery backups to keep alarms running during outages. Utilizing hydrocarbon refrigerants, this unit meets EPA SNAP compliance, protecting environmental safety. It features a durable white powder-coated steel exterior, with a keyed door lock for added security. This unit offers 30 cubic feet of spacious storage and five adjustable shelves for versatile organization. With a low energy consumption and environmental benefits, the ABT-HC-30R-P is an efficient and next-level solution for diverse laboratory applications.

In stock
SKU
CESS-876725-00001
MPN
ABT-HC-30R-P
$10,678.20
Specifications
DiscontinuedNo
Additional Specifications:
Storage capacity (cu. ft) 30 Cu. Ft.
Door One solid swing, right hinged; stainless steel door handle; magnetic door gasket
Shelves Five adjustable
Mounting and Installation Caster platform
Airflow Management Forced draft circulation
External probe access 3/4" probe access port on rear wall
Insulation EPA compliant high density urethane foam
Exterior materials Powder coated steel
Access Control Keyed door lock
General Warranty Two (2) year parts and labor warranty
Compressor Warranty Five (5) year compressor parts warranty
Product Weight (Lbs) 350
Shipping Weight (Lbs) 380
Rated Amperage 3.1 Amps
Power Plug/Power Cord NEMA 5-15P plug, 6 ft long cord
Facility Electrical Requirement 110 - 120V AC, 15A breaker, NEMA 5-15 receptacle
Agency and Certification UL listed per UL 60335-1 and UL 60335-2-89 standards; c UL listed per CSA C22.2
Uniformity (Cabinet air) +/- 1.6°C
Stability (Cabinet air) +/- 1.0°C
Maximum temperature variation
(Cabinet air)
3.6 °C Total
Energy Consumption (KWh/day) < 1.7 KWh/day
Average Heat Rejection (BTU/hr) 520 BTU/hr
Pull down time to nominal operating temp <35 min (empty)
Compressor Hermetic, variable speed
Refrigerant EPA SNAP compliant R600a Isobutane
Condenser Static - exterior walls of cabinet
Evaporator Fin and tube design, high efficiency fan
Defrost Off-cycle defrost, no heat added
Controller technology Microprocessor temperature controller with variable frequency compressor control
Battery Backup Optional accessory
Temperature Setpoint Range Notes Minimum temperature limited to avoid freezing
Adjustable Temperature Range 2°C to 10°C
External alarm connection Remote alarm contacts
Alarms Audible and visual high/low-temperature alarms, Door ajar alarm, Remote alarm contacts
Display Probe Product simulator bottle with glass bead media
Disclaimers Refrigerator requires a minimum 4” of clearance around back and sides of the unit for proper ventilation (more if next to heat rejecting equipment) - improper installation will void the warranty and will lead to temperature maintenance issues, overheating and possible failure