These premier undercounter refrigerators are designed in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste.
These glass door freestanding refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, LED interior lighting, and probe access ports with included probes. American Biotech Supply vaccine storage refrigerators utilize HFC-free refrigerant for environmental health and energy efficiency.
For more information on Vaccine Storage: Vaccine Storage
The acceptable temperature ranges for vaccine storage are 36?F to 46?F (or 2?C and 8?C) in the refrigerator and -58?F and 5?F (or -50?C and -15?C) in freezers.
Features:
Operational environment: Indoor use only. Optimal operating range: +18?C to +26?C (+65?F to +78?F), <70% RH
Door: One swing glass door, self-closing, right hinged, non-reversible, magnetic sealed gasket, keyed lock
Shelves: Four shelves (three adjustable/one fixed) with guard rail on back
Mounting and Installation: Leveling legs. Note: 4" of clearance on all sides must be maintained for adequate ventilation
Interior lighting: Shielded, switched LED lighting, full coverage, balanced spectrum
Airflow management: Forced Air technology, patent pending
External probe access: Rear wall port (0.38") dia.
Exterior materials: White powder coated steel
Access control: Pyxis, Omnicell and AcuDose RX compatible
Power Plug/Power Cord: NEMA 5-15 plug, 8 to 10 ft typical, conforms to UL471 requirements, Vaccine storage power cord warning label
Facility Electrical Requirement: 110-120V AC: 15 A (minimum)
Agency Listing and Certification: Certified in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. UL, C-UL, ETL, C-ETL listed (either single or dual agency listings) and certified to UL471 standard, hydrocarbon refrigerant safety.
Included Accessories: Temperature monitor device (TMD) complies with the current CDC guidelines, with 3 years certification of calibration, "buffered" probe in the product simulated solution, min/max memory, field installable, and visual & audible temp alarm. Pharmacy refrigerator/freezer toolkit and temperature logs
Performance:
Uniformity (Cabinet air): +/- 1.1?C
Stability (Cabinet air): +/- 0.7?C
Maximum temperature variation: +/-1.2?C
Temperature rise after 8 sec door openings: Temperature did not exceed 7.2?C at any probe for all required NSF/ANSI 456 testing protocols
Recovery after 3 min door opening: All probes recover to under 8?C within 7.5 min.
Energy consumption: 0.58 KWh/day
Average heat rejection: 1.00 KWh/day (142 BTU/h)
Noise pressure level (dBA): 34 or less installed
Pull down time to nominal operating temp: 42 min
Notes:
Performance data acquired at 22?C ambient, using NSF/ANSI 456 compliant validation ballast probes, empty chamber, during stabilized steady state operation and a DAQ sampling rate of one measurement every 10 seconds.
Uniformity is defined as the maximum variance in temperature across all probes at any point in time over the testing period
Stability is defined as the maximum variance in temperature experienced by any single probe over the testing period
Temperature performance for all loaded and unloaded door opening protocols, all alarm, controller and probe requirements as defined in the NSF/ANSI 456 standard for vaccine storage
Data per Energy Star test results or equivalent testing and calculation. Heat rejection based on daily averages, not continuous operation. Performance exceeds Energy Star requirements.
Refrigeration System
Compressor: Hermetic, high performance
Refrigerant: EPA SNAP compliant, R600a, Isobutane
Condenser: Tube and grid construction, fanless
Evaporator: Plate wall
Defrost: Cycle optimized, zero energy
Temp Controller:
Controller technology: Parametric, microprocessor, LED display with 0.1?C resolution
Temperature setpoint range: 1?C to 10?C (Setpoint must remain unaltered from the factory setting to remain compliant with NSF/ANSI 456 Standard for Vaccine Storage requirements)
Display probe: Calibrated, stainless steel
External alarm connection: State switching remote alarm contacts
Alarms: Visual and audible indicators. High / Low temperature, compliant with alarm requirements defined in the NSF/ANSI 456 Standard for Vaccine Storage
Simulator ballast: Glass bead thermal media
Quote Information
Commercial Procurement Made Easier
New healthcare facility construction, expansion or remodel - we’ve created a
streamlined approach to healthcare supply chain for our following customers:
Medical facilities (Hospitals, Physician Offices, Long Term Care facilities,
etc.)
GPO members (Vizient, Premier, Intalere, Healthrust, ROi)
Simply search for your desired product(s), and click ‘Chat with Us for GPO Pricing!’
to connect with one of our product specialists. They will note your contact
information and any comments/questions about the item or an overall quote,
and submit it to an account manager. One of our dedicated account managers
will reach out to you with your pricing and to discuss your logistics needs
within 12 hours.
CME Makes It Easy
2 million+ products, from over 2000 manufacturers.
Contract pricing with all the major GPOs, and volume discounts for bulk orders.
We save you time and money by being your most competitive source for all items
in your PO/quote, and we’ll even locate those products we don’t carry
at the best price.
Streamlined, on-time delivery with turn-key white glove service, scheduled for
when its best for you.
Quote to Delivery. One Team. One Partner. CME Makes It Easy.